Can J Cardiol. 2015 Sep;31(9):1204.e13-5. doi: 10.1016/j.cjca.2015.02.011. Epub 2015 Feb 17.

Emergency Transcatheter Aortic Valve Implantation for Acute and Early Failure of Sutureless Perceval Aortic Valve.

Author information

1
Department of Cardiology, Rouen University Hospital, INSERM U1096, Rouen, France. Electronic address: eric.durand@chu-rouen.fr.
2
Department of Cardiology, Rouen University Hospital, INSERM U1096, Rouen, France.

Abstract

We report the case of a 78-year-old woman admitted for cardiogenic shock related to acute and early failure (severe aortic regurgitation) of a Perceval sutureless aortic bioprosthesis (Sorin Group, Saluggia, Italy). Clinical stability was achieved using rescue transfemoral transcatheter aortic valve-in-valve implantation with an Edwards SAPIEN 3 prosthesis (Edwards Lifesciences, Irvine, CA). To our knowledge, we report herein the first case of successful valve-in-valve implantation using a SAPIEN 3 transcatheter heart valve in a sutureless bioprosthetic aortic valve with acute and early deterioration.