Radiology. 2015 Aug;276(2):465-78. doi: 10.1148/radiol.2015141991. Epub 2015 May 26.

Observer Performance in the Detection and Classification of Malignant Hepatic Nodules and Masses with CT Image-Space Denoising and Iterative Reconstruction.

Author information

1
From the Departments of Radiology (J.G.F., L.Y., Z.L., D.M.H., S.K.V., J.L.F., M.S., D.L., S.L., C.H.M.), Physiology and Biomedical Engineering (A.M., D.S.L., K.E.A., D.R.H.), Information Technology (D.J.B.), and Biomedical Statistics and Informatics (R.E.C.), Mayo Clinic, 200 First St SW, Rochester, MN 55905; Department of Radiology, Mayo Clinic, Eau Claire, Wis (G.C.B.); Department of Radiology, Mayo Clinic, Jacksonville, Fla (J.C.C.); and Department of Radiology, Mayo Clinic, Scottsdale, Ariz (A.K.H.).

Abstract

PURPOSE:

To determine if lower-dose computed tomographic (CT) scans obtained with adaptive image-based noise reduction (adaptive nonlocal means [ANLM]) or iterative reconstruction (sinogram-affirmed iterative reconstruction [SAFIRE]) result in reduced observer performance in the detection of malignant hepatic nodules and masses compared with routine-dose scans obtained with filtered back projection (FBP).

MATERIALS AND METHODS:

This study was approved by the institutional review board and was compliant with HIPAA. Informed consent was obtained from patients for the retrospective use of medical records for research purposes. CT projection data from 33 abdominal and 27 liver or pancreas CT examinations were collected (median volume CT dose index, 13.8 and 24.0 mGy, respectively). Hepatic malignancy was defined by progression or regression or with histopathologic findings. Lower-dose data were created by using a validated noise insertion method (10.4 mGy for abdominal CT and 14.6 mGy for liver or pancreas CT) and images reconstructed with FBP, ANLM, and SAFIRE. Four readers evaluated routine-dose FBP images and all lower-dose images, circumscribing liver lesions and selecting diagnosis. The jackknife free-response receiver operating characteristic figure of merit (FOM) was calculated on a per-malignant nodule or per-mass basis. Noninferiority was defined by the lower limit of the 95% confidence interval (CI) of the difference between lower-dose and routine-dose FOMs being less than -0.10.

RESULTS:

Twenty-nine patients had 62 malignant hepatic nodules and masses. Estimated FOM differences between lower-dose FBP and lower-dose ANLM versus routine-dose FBP were noninferior (difference: -0.041 [95% CI: -0.090, 0.009] and -0.003 [95% CI: -0.052, 0.047], respectively). In patients with dedicated liver scans, lower-dose ANLM images were noninferior (difference: +0.015 [95% CI: -0.077, 0.106]), whereas lower-dose FBP images were not (difference -0.049 [95% CI: -0.140, 0.043]). In 37 patients with SAFIRE reconstructions, the three lower-dose alternatives were found to be noninferior to the routine-dose FBP.

CONCLUSION:

At moderate levels of dose reduction, lower-dose FBP images without ANLM or SAFIRE were noninferior to routine-dose images for abdominal CT but not for liver or pancreas CT.

PMID:
26020436
PMCID:
PMC4514571
DOI:
10.1148/radiol.2015141991
[Indexed for MEDLINE]