Biomed Tech (Berl). 2015 Dec;60(6):541-9. doi: 10.1515/bmt-2014-0184.

Measurement of susceptibility artifacts with histogram-based reference value on magnetic resonance images according to standard ASTM F2119.

Abstract

OBJECTIVE:

The standard ASTM F2119 describes a test method for measuring the size of a susceptibility artifact based on the example of a passive implant. A pixel in an image is considered to be a part of an image artifact if the intensity is changed by at least 30% in the presence of a test object, compared to a reference image in which the test object is absent (reference value). The aim of this paper is to simplify and accelerate the test method using a histogram-based reference value.

MATERIALS AND METHODS:

Four test objects were scanned parallel and perpendicular to the main magnetic field, and the largest susceptibility artifacts were measured using two methods of reference value determination (reference image-based and histogram-based reference value). The results between both methods were compared using the Mann-Whitney U-test.

RESULTS:

The difference between both reference values was 42.35 ± 23.66. The difference of artifact size was 0.64 ± 0.69 mm. The artifact sizes of both methods did not show significant differences; the p-value of the Mann-Whitney U-test was between 0.710 and 0.521.

CONCLUSIONS:

A standard-conform method for a rapid, objective, and reproducible evaluation of susceptibility artifacts could be implemented. The result of the histogram-based method does not significantly differ from the ASTM-conform method.

PMID:
25992509
DOI:
10.1515/bmt-2014-0184
[Indexed for MEDLINE]