Improving the detection of small lesions using a state-of-the-art time-of-flight PET/CT system and small-voxel reconstructions

J Nucl Med Technol. 2015 Mar;43(1):21-7. doi: 10.2967/jnmt.114.147215. Epub 2015 Jan 22.

Abstract

A major disadvantage of (18)F-FDG PET involves poor detection of small lesions and lesions with low metabolism, caused by limited spatial resolution and relatively large image voxel size. As spatial resolution and sensitivity are better in new PET systems, it is expected that small-lesion detection could be improved using smaller voxels. The aim of this study was to test this hypothesis using a state-of-the-art time-of-flight PET/CT device.

Methods: (18)F-FDG PET scans of 2 image-quality phantoms (sphere sizes, 4-37 mm) and 39 consecutive patients with lung cancer were analyzed on a time-of-flight PET/CT system. Images were iteratively reconstructed with standard 4 × 4 × 4 mm voxels and smaller 2 × 2 × 2 mm voxels. For the phantom study, we determined contrast-recovery coefficients and signal-to-noise ratios (SNRs). For the patient study, (18)F-FDG PET-positive lesions in the chest and upper abdomen with a volume less than 3.0 mL (diameter, <18 mm) were included. Lesion mean and maximum standardized uptake values (SUVmean and SUVmax, respectively) were determined in both image sets. SNRs were determined by comparing SUVmax and SUVmean with background noise levels. A subanalysis was performed for lesions less than 0.75 mL (diameter, <11 mm). For qualitative analysis of patient data, 3 experienced nuclear medicine physicians gave their preference after visual side-by-side analysis.

Results: For phantom spheres 13 mm or less, we found higher contrast-recovery coefficients and SNRs using small-voxel reconstructions. For 66 included (18)F-FDG PET-positive lesions, the average increase in SUVmean and SUVmax using the small-voxel images was 17% and 32%, respectively (P < 0.01). For lesions less than 0.75 mL (21 in total), the average increase was 21% and 44%, respectively. Moreover, averaged over all lesions, the mean and maximum SNR increased by 20% and 27%, respectively (P < 0.01). For lesions less than 0.75 mL, these values increased up to 23% and 46%, respectively. The physicians preferred the small-voxel reconstructions in 76% of cases.

Conclusion: Supported by a phantom study, there was a visual preference toward (18)F-FDG PET images reconstructed with 2 × 2 × 2 mm voxels and a profound increase in standardized uptake value and SNR for small lesions. Hence, it is expected that small-lesion detection improves using small-voxel reconstructions.

Keywords: 18F-FDG PET; lung cancer; small lesions; small-voxel reconstruction; time-of-flight.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Fluorodeoxyglucose F18
  • Humans
  • Image Processing, Computer-Assisted / methods*
  • Lung Neoplasms / diagnosis*
  • Lung Neoplasms / diagnostic imaging
  • Lung Neoplasms / pathology*
  • Multimodal Imaging / instrumentation
  • Multimodal Imaging / methods*
  • Phantoms, Imaging
  • Positron-Emission Tomography / instrumentation
  • Positron-Emission Tomography / methods*
  • Time Factors
  • Tomography, X-Ray Computed / instrumentation
  • Tomography, X-Ray Computed / methods*
  • Tumor Burden*