Hematopoietic progenitor cells (HPC) are stored in cryopreservation bags that are resistant to liquid nitrogen. Since Cryocyte bags of Baxter (B-bags) are no longer available, an alternative bag was sought. Also, the influence of freezing volume was studied. Miltenyi Biotec (MB)- and MacoPharma (MP)-bags passed the integrity tests without failure. Comparing MB- and MP-bags with B-bags, no difference in WBC recovery or viability was found when using a WBC-enriched product as a "dummy" HPC product. Further, a freezing volume of 30 mL resulted in better WBC recovery and viability than 60 mL. Additonal studies using real HPC might be necessary.
Keywords: Cryopreservation bags; Integrity; WBC recovery; WBC viability; WBC-enriched product.
Copyright © 2014 Elsevier Ltd. All rights reserved.