Morphological and electrochemical properties of an explanted PtIr electrode array after 15 months in vivo

Annu Int Conf IEEE Eng Med Biol Soc. 2014:2014:418-21. doi: 10.1109/EMBC.2014.6943617.

Abstract

We investigated the morphological and electrochemical properties of an explanted laser-machined 32 channel electrocorticogram (ECoG) electrode array made of platinum-iridium and silicone rubber. It was connected to a wireless brain-computer interface (BCI) and implanted in a sheep for more than 15 months. Recordings and stimulations of cortical activity were conducted over the whole period on a regularly basis. Currently, this is the longest in vivo study for this type of ECoG electrode array. Results were compared with an unused electrode array of same dimensions, material and production method. Visual inspections revealed no significant material alterations, despite organic residuals which could be easily removed though. Electrochemical impedance measurements also attested proper long-term stability of magnitude and phase, the difference between explanted electrode contacts and those of the unused array were found negligible.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Brain-Computer Interfaces
  • Dielectric Spectroscopy
  • Electrochemistry*
  • Electrodes, Implanted*
  • Electroencephalography
  • Iridium / chemistry*
  • Microscopy, Electron, Scanning
  • Platinum / chemistry*
  • Sheep