Format

Send to

Choose Destination
Doc Ophthalmol. 2015 Feb;130(1):1-12. doi: 10.1007/s10633-014-9473-7. Epub 2014 Dec 14.

ISCEV Standard for full-field clinical electroretinography (2015 update).

Author information

1
School of Optometry and Vision Sciences, University of Waterloo, 200 University Avenue West, Waterloo, ON, N2L 3G1, Canada, daphne.mcculloch@uwaterloo.ca.

Abstract

This document, from the International Society for Clinical Electrophysiology of Vision (ISCEV), presents an updated and revised ISCEV Standard for full-field clinical electroretinography (ffERG or simply ERG). The parameters for Standard flash stimuli have been revised to accommodate a variety of light sources including gas discharge lamps and light emitting diodes. This ISCEV Standard for clinical ERGs specifies six responses based on the adaptation state of the eye and the flash strength: (1) Dark-adapted 0.01 ERG (rod ERG); (2) Dark-adapted 3 ERG (combined rod-cone standard flash ERG); (3) Dark-adapted 3 oscillatory potentials; (4) Dark-adapted 10 ERG (strong flash ERG); (5) Light-adapted 3 ERG (standard flash "cone" ERG); and (6) Light-adapted 30 Hz flicker ERG. ISCEV encourages the use of additional ERG protocols for testing beyond this minimum standard for clinical ERGs.

PMID:
25502644
DOI:
10.1007/s10633-014-9473-7
[Indexed for MEDLINE]

Supplemental Content

Full text links

Icon for Springer
Loading ...
Support Center