Magn Reson Med. 2015 Dec;74(6):1621-31. doi: 10.1002/mrm.25559. Epub 2014 Dec 9.

MR fingerprinting using fast imaging with steady state precession (FISP) with spiral readout.

Author information

1
Department of Biomedical Engineering, Case Western Reserve University, Cleveland, Ohio, USA.
2
Department of Radiology, University Hospitals Case Medical Center, Case Western Reserve University, Cleveland, Ohio, USA.

Abstract

PURPOSE:

This study explores the possibility of using gradient echo-based sequences other than balanced steady-state free precession (bSSFP) in the magnetic resonance fingerprinting (MRF) framework to quantify the relaxation parameters .

METHODS:

An MRF method based on a fast imaging with steady-state precession (FISP) sequence structure is presented. A dictionary containing possible signal evolutions with physiological range of T1 and T2 was created using the extended phase graph formalism according to the acquisition parameters. The proposed method was evaluated in a phantom and a human brain. T1 , T2 , and proton density were quantified directly from the undersampled data by the pattern recognition algorithm.

RESULTS:

T1 and T2 values from the phantom demonstrate that the results of MRF FISP are in good agreement with the traditional gold-standard methods. T1 and T2 values in brain are within the range of previously reported values.

CONCLUSION:

MRF-FISP enables a fast and accurate quantification of the relaxation parameters. It is immune to the banding artifact of bSSFP due to B0 inhomogeneities, which could improve the ability to use MRF for applications beyond brain imaging.

KEYWORDS:

FISP; MR fingerprinting; quantitative imaging; relaxation time; spiral

PMID:
25491018
PMCID:
PMC4461545
DOI:
10.1002/mrm.25559
[Indexed for MEDLINE]