Am J Perinatol. 2015 Jun;32(7):639-44. doi: 10.1055/s-0034-1390345. Epub 2014 Oct 26.

Factors Influencing the Accuracy of Noninvasive Blood Pressure Measurements in NICU Infants.

Author information

1
Division of Neonatal-Perinatal Medicine, Department of Pediatrics, Baylor College of Medicine, Houston, Texas.
2
Division of Neonatal-Perinatal Medicine, Department of Pediatrics, University of North Carolina, Chapel Hill, North Carolina.

Abstract

OBJECTIVE:

Compare invasive blood pressure (IBP) and noninvasive blood pressure (NIBP) measurement methods in the neonatal intensive care unit (NICU) across various gestational age and postmenstrual age (PMA), and determine the effect of gestational age and PMA on accuracy of NIBP measurements.

STUDY DESIGN:

Retrospective chart review of paired mean IBP and NIBP measurements from infants admitted to a single NICU from January 2008 through December 2010. Infants with congenital anomalies or receiving therapeutic hypothermia were excluded. Difference between paired measurements was analyzed using Bland-Altman method. We examined the association between PMA, sex, race, mechanical ventilation, medications, and axillary temperature, and the difference in measurements using a mixed effects linear regression model.

RESULTS:

Eighty-seven infants had 243 observations. The mean (range) gestational age at birth was 31.9 (23-41) weeks and PMA at time of measurement ranged from 26 to 52 weeks. We found poor agreement between IBP and NIBP measurements, with mean difference (95% limits of agreement) of -8.8 (11, -28.7) mm Hg. The mean blood pressure percent difference ( ±  SD) was -28.3 ( ±  35.6%). A greater blood pressure percent difference between the two measurement techniques was associated with lower PMA and lower mean IBP.

CONCLUSION:

NIBP measurements overestimate IBP measurements particularly in smaller, sicker infants at lower IBP measurements.

PMID:
25344873
DOI:
10.1055/s-0034-1390345
[Indexed for MEDLINE]