Format

Send to

Choose Destination
Graefes Arch Clin Exp Ophthalmol. 2015 Jun;253(6):849-54. doi: 10.1007/s00417-014-2774-4. Epub 2014 Aug 27.

Factors affecting laser power in retinal Navilas laser treatment.

Author information

1
Department of Ophthalmology, University Hospital Heidelberg, Im Neuenheimer Feld 400, 69120, Heidelberg, Germany, alexandra.hoeh@med.uni-heidelberg.de.

Abstract

PURPOSE:

To evaluate the effect of patient-associated factors on the minimum laser power needed for a mild visible burn in focal laser treatments using the 532 nm Navilas laser system.

METHODS:

We conducted a monocentric prospective pilot study of 58 eyes of 40 patients with diabetic macular edema. The following parameters were analysed: axial length, refraction, iris pigmentation, lens status, lens grading and densitometry, retinal and choroidal thickness and focus setting during treatment. Laser power was adjusted to produce mild, barely visible burns. Retinal laser burn size was measured 30 min after treatment.

RESULTS:

Focus setting is significantly correlated with retinal lesion size (r = 0.50, p = 0.001) and laser power (r = 0.44, p < 0.001). Axial length only correlated with laser power when the effect of focus was controlled. Phakic eyes needed more laser power than pseudophakic eyes (78.3 versus 67.2 mW, p = 0.051). No correlation of laser power with any other factor could be found.

CONCLUSIONS:

Among the examined parameters, focus setting had the strongest effect on the laser power needed to produce a mild visible burn. The association of focus with laser power can be explained by the focus-dependent change of retinal spot size. Lens status (phakic versus pseudophakic patients) seems to influence laser light transmission in the examined age group.

PMID:
25161075
DOI:
10.1007/s00417-014-2774-4
[Indexed for MEDLINE]