Br J Radiol. 2014 Sep;87(1041):20140110. doi: 10.1259/bjr.20140110. Epub 2014 Jul 7.

Comparison of effective dose and lifetime risk of cancer incidence of CT attenuation correction acquisitions and radiopharmaceutical administration for myocardial perfusion imaging.

Author information

1
The University of Salford, Directorate of Radiography, Salford, Greater Manchester, UK.

Abstract

OBJECTIVE:

To measure the organ dose and calculate effective dose from CT attenuation correction (CTAC) acquisitions from four commonly used gamma camera single photon emission CT/CT systems.

METHODS:

CTAC dosimetry data was collected using thermoluminescent dosemeters on GE Healthcare's Infinia™ Hawkeye™ (GE Healthcare, Buckinghamshire, UK) four- and single-slice systems, Siemens Symbia™ T6 (Siemens Healthcare, Erlangen, Germany) and the Philips Precedence (Philips Healthcare, Amsterdam, Netherlands). Organ and effective dose from the administration of (99m)Tc-tetrofosmin and (99m)Tc-sestamibi were calculated using International Commission of Radiological Protection reports 80 and 106. Using these data, the lifetime biological risk was calculated.

RESULTS:

The Siemens Symbia gave the lowest CTAC dose (1.8 mSv) followed by the GE Infinia Hawkeye single-slice (1.9 mSv), GE Infinia Hawkeye four-slice (2.5 mSv) and Philips Precedence v. 3.0. Doses were significantly lower than the calculated doses from radiopharmaceutical administration (11 and 14 mSv for (99m)Tc-tetrofosmin and (99m)Tc-sestamibi, respectively). Overall lifetime biological risks were lower, which suggests that using CTAC data posed minimal risk to the patient. Comparison of data for breast tissue demonstrated a higher risk than that from the radiopharmaceutical administration.

CONCLUSION:

CTAC doses were confirmed to be much lower than those from radiopharmaceutical administration. The localized nature of the CTAC exposure compared to the radiopharmaceutical biological distribution indicated dose and risk to the breast to be higher.

ADVANCES IN KNOWLEDGE:

This research proved that CTAC is a comparatively low-dose acquisition. However, it has been shown that there is increased risk for breast tissue especially in the younger patients. As per legislation, justification is required and CTAC should only be used in situations that demonstrate sufficient net benefit.

PMID:
24998249
PMCID:
PMC4453143
DOI:
10.1259/bjr.20140110
[Indexed for MEDLINE]