Stainless steel wire versus FiberWire suture cerclage fixation to stabilize the humerus in total shoulder arthroplasty.

Author information

1
Department of Orthopaedics, University of Zürich, Balgrist University Hospital, Zürich, Switzerland. Electronic address: niklas.renner@balgrist.ch.
2
Department of Orthopaedics, University of Zürich, Balgrist University Hospital, Zürich, Switzerland.
3
Department of Ortrhopaedics, Private Hospital Maria Hilf, Klagenfurt, Austria.

Abstract

HYPOTHESIS:

No. 5 FiberWire (Arthrex, Naples, FL, USA) cerclage (FWC) and 1.25-mm stainless steel wire cerclage (SSWC) are biomechanically similar in resistance to prosthetic subsidence in shoulder arthroplasty.

METHODS:

In this laboratory bench study, 3 different surgical knot configurations (4-throw knot, cow hitch, and simple hitch) using a No. 5 FWC were evaluated and compared with a 1.25-mm SSWC. First, distraction tests were performed using bovine femoral cortical half shells mounted on a testing jig. Cerclage tightening, load to a 3-mm gap opening, and load to total failure were measured. Second, uncemented humeral prosthetic stems were inserted into an experimentally split humeral medullary canal, secured by the cerclage. After 100 N of preloading, the prosthesis was advanced into the humerus at a speed of 0.2 mm/s, and resistance during subsidence up to a penetration depth of 10 mm, as well as gap opening, was measured.

RESULTS:

Tightening force showed higher values for SSWC (618 N) than FWC (131-137 N) (P < .001). Load to total failure was comparable among the 3 different FWC knots (2,642-2,804 N), which were significantly stronger than SSWC (1,775 N, P < .001). At 3 mm of distraction, SSWC (1,820 N), cow hitch (1,803 N), and single-throw hitch (1,709 N) performed significantly better than a 4-throw knot (1,289 N) (P < .01). Subsidence testing showed no difference in force restraint or gap opening between the best FWC and SSWC.

CONCLUSIONS:

FWCs appear, in vitro, equally suitable to steel wires to stabilize nondisplaced periprosthetic humeral fractures. To actively reduce a displaced fracture, steel wires may still be the first choice.

KEYWORDS:

Cerclage; periprosthetic humeral fracture; total shoulder arthroplasty