Acta Radiol. 2015 Apr;56(4):471-6. doi: 10.1177/0284185114531572. Epub 2014 Apr 29.

Effectiveness of a silicone device for foot MRI in order to obtain homogeneous fat suppression images.

Author information

1
Department of Radiology, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Republic of Korea.
2
Department of Radiological Science, College of Health Sciences, Eulji University, Gyeonggi-do, Republic of Korea handk@eulji.ac.kr.

Abstract

BACKGROUND:

Incomplete fat suppression induced by magnetic field inhomogeneity is difficult to compensate for with hardware magnetic-field shimming.

PURPOSE:

To evaluate the effectiveness of a silicone device used to obtain homogeneous fat suppression during 3T magnetic resonance imaging (MRI) scans of the foot.

MATERIAL AND METHODS:

Thirty-eight healthy volunteers were enrolled and examined twice, before (group A) and after (group B) the application of a silicone device. Fat-saturated, T2-weighted, fast spin-echo images were acquired using the same scanning protocol at both examinations. Signal- and contrast-to-noise ratios (SNR and CNR) were calculated and compared in the four regions of interest (ROIs). ROI 1 and 2 were selected from toe-side bone and soft tissue, while ROI 3 and 4 were selected from proximal bone and soft tissue. Qualitative analysis using a four-point scale was performed for three categories. The categories are as follows: the overall image quality, homogeneity of the first phalange and metatarsal bone, respectively.

RESULTS:

The SNR and CNR in ROI 1 and 2 were significantly higher in group A than in group B (SNR; P < 0.001, CNR; P < 0.001), and there were no significant difference in ROI 3 and 4. The qualitative score of the overall fat suppression in group B was significantly higher than that in group A (P < 0.001). Homogeneity of the first phalange in group B was also significantly higher than that in group A (P < 0.001). On the other hand, the homogeneity of the metatarsal bone was not significantly different in the two groups.

CONCLUSION:

The use of a silicone device provides homogeneous fat suppression in 3T MRI of the foot and can significantly improve image quality.

KEYWORDS:

Fat suppression; contrast-to-noise ratio; homogeneity; magnetic resonance imaging; signal-to-noise ratio; silicone

PMID:
24782573
DOI:
10.1177/0284185114531572
[Indexed for MEDLINE]