Format

Send to

Choose Destination
Neurointervention. 2014 Feb;9(1):21-5. doi: 10.5469/neuroint.2014.9.1.21. Epub 2014 Feb 28.

Manual Aspiration Technique to Retrieve a Prematurely Detached Coil during Cerebral Aneurysm Embolization.

Author information

1
Department of Neurosurgery, Rush University Medical Center, Chicago, IL, USA.

Abstract

Coil herniation, premature deployment, and fracture are procedure associated complications of endovascular aneurysm embolization that optimally necessitate coil retrieval when feasible. Several published techniques describe different strategies for managing coil complications including various snare retrieval devices, alligator retrieval devices, stent fixation, and open surgical resection of coils when distal blood flow is compromised. We report a novel technique employed to retrieve a prematurely detached coil during an aneurysm embolization using a syringe fixed to the microcatheter to carefully aspirate a loose coil with direct fluoroscopic visualization. This technique can only be utilized in the circumstance where the proximal end of the coil remains in the microcatheter. Conventional techniques of coil retrieval and stenting are discussed and compared to the rational for using the manual aspiration technique.

KEYWORDS:

Cerebral aneuryusms; Embolization, therapeutic; Endovascular procedures/complication

Supplemental Content

Full text links

Icon for Publishing M2Community Icon for PubMed Central
Loading ...
Support Center