Sensors (Basel). 2013 Dec 13;13(12):17281-91. doi: 10.3390/s131217281.

Reliability of measured data for pH sensor arrays with fault diagnosis and data fusion based on LabVIEW.

Author information

1
Department of Information Management, Transworld University, 1221 Zhennan Rd., Yunlin 64063, Taiwan. liaoih@twu.edu.tw.

Abstract

Fault diagnosis (FD) and data fusion (DF) technologies implemented in the LabVIEW program were used for a ruthenium dioxide pH sensor array. The purpose of the fault diagnosis and data fusion technologies is to increase the reliability of measured data. Data fusion is a very useful statistical method used for sensor arrays in many fields. Fault diagnosis is used to avoid sensor faults and to measure errors in the electrochemical measurement system, therefore, in this study, we use fault diagnosis to remove any faulty sensors in advance, and then proceed with data fusion in the sensor array. The average, self-adaptive and coefficient of variance data fusion methods are used in this study. The pH electrode is fabricated with ruthenium dioxide (RuO2) sensing membrane using a sputtering system to deposit it onto a silicon substrate, and eight RuO2 pH electrodes are fabricated to form a sensor array for this study.