Novel tandem quadrupole-acceleration-deceleration mass spectrometer for neutralization-reionization studies

J Am Soc Mass Spectrom. 1992 Jul;3(5):493-501. doi: 10.1016/1044-0305(92)85026-G.

Abstract

A new tandem mass spectrometer of the quadrupole-acceleration lens-deceleration. lens-quadrupole (QADQ) configuration is described. The instrument is designed for neutralization-reionization studies and consists of a 2000-u quadrupole mass analyzer as MS-I, an acceleration electrostatic lens, a series of three differentially pumped collision cells, and an electrostatic deceleration lens, energy filter, and another 2000-u quadrupole mass analyzer as MS-II. The ion optical system achieves high total ion transmission for 5-9-keV ions. Unit mass resolution in neutralization-reionization mass spectra of aromatic compounds is demonstrated. Mass, kinetic energy, and linked scans at various levels of mass resolution and sensitivity are described.