J Allergy Clin Immunol Pract. 2013 Jan;1(1):83-91. doi: 10.1016/j.jaip.2012.08.004. Epub 2012 Nov 2.

Use of metered-dose inhaler electronic monitoring in a real-world asthma randomized controlled trial.

Author information

1
Medical Research Institute of New Zealand, Wellington, New Zealand; Capital & Coast District Health Board, Wellington, New Zealand; Division of Respiratory Medicine, School of Clinical Sciences, University of Nottingham, Nottingham, United Kingdom. Electronic address: mitesh.patel@mrinz.ac.nz.

Erratum in

  • J Allergy Clin Immunol Pract. 2013 May-Jun;1(3):314.

Abstract

BACKGROUND:

Electronic monitoring of inhaled asthma medications is one method to measure medication adherence and patterns of use. Information on the performance of monitors in a randomized controlled trial allows researchers and clinicians to understand their utility and limitations. The Smartinhaler Tracker is an electronic monitor for metered-dose inhalers (MDIs) that records the date, time, and number of actuations.

OBJECTIVE:

To determine the performance of the Smartinhaler monitors used in a 24-week randomized controlled trial of 303 patients with asthma in a real-world setting.

METHODS:

Prestudy use checks involved 2 actuations of the MDI, with a further 2 performed 2 hours later. Within-study monitor checks, performed before dispensing at clinic visits 2 to 4, included a computerized check of monitor clock function, actuation accuracy, and battery life. Within-study data checks involved computerized checks of monitor clock function before data upload.

RESULTS:

Two thousand six hundred seventy-eight of 2728 monitors (98.2%) passed prestudy use checks. Seventy-six of 2642 monitors (2.9%) dispensed to participants failed within-study monitor checks. Fifty-one of 2642 monitors (1.9%) malfunctioned before data upload, mostly as a result of fluid immersion. Ninety-three of 2642 monitors (3.5%) were lost or thrown away by participants. Complete data was available from 2498 of 2642 dispensed monitors (94.5%) and 2498 of 2549 returned monitors (98.0%).

CONCLUSIONS:

The Smartinhaler Tracker is a reliable monitor for measuring MDI use in a real-world setting. Use of extensive monitor and data-checking protocols reduces data loss. In a research or clinical setting, the use of a validated and reliable electronic monitor represents the reference standard for assessing patterns of medication use.

KEYWORDS:

Adherence; Asthma; Electronic; Fixed-dose combination inhaler as maintenance therapy with short-acting beta agonist as reliever; MDI; Metered-dose inhaler; Monitoring; Patterns of use; RCT; Randomized controlled trial; SMART; STANDARD; Single combination inhaler as Maintenance And Reliever Therapy; Smartinhaler; Validation

PMID:
24229826
DOI:
10.1016/j.jaip.2012.08.004
[Indexed for MEDLINE]
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for Elsevier Science
    Loading ...
    Support Center