Diabetes Technol Ther. 2014 Feb;16(2):97-101. doi: 10.1089/dia.2013.0221. Epub 2013 Nov 4.

Accuracy of subcutaneous continuous glucose monitoring in critically ill adults: improved sensor performance with enhanced calibrations.

Author information

1
1 Metabolic Research Laboratories, Institute of Metabolic Science, University of Cambridge , Cambridge, United Kingdom .

Abstract

OBJECTIVE:

Accurate real-time continuous glucose measurements may improve glucose control in the critical care unit. We evaluated the accuracy of the FreeStyle(®) Navigator(®) (Abbott Diabetes Care, Alameda, CA) subcutaneous continuous glucose monitoring (CGM) device in critically ill adults using two methods of calibration.

SUBJECTS AND METHODS:

In a randomized trial, paired CGM and reference glucose (hourly arterial blood glucose [ABG]) were collected over a 48-h period from 24 adults with critical illness (mean±SD age, 60±14 years; mean±SD body mass index, 29.6±9.3 kg/m(2); mean±SD Acute Physiology and Chronic Health Evaluation score, 12±4 [range, 6-19]) and hyperglycemia. In 12 subjects, the CGM device was calibrated at variable intervals of 1-6 h using ABG. In the other 12 subjects, the sensor was calibrated according to the manufacturer's instructions (1, 2, 10, and 24 h) using arterial blood and the built-in point-of-care glucometer.

RESULTS:

In total, 1,060 CGM-ABG pairs were analyzed over the glucose range from 4.3 to 18.8 mmol/L. Using enhanced calibration median (interquartile range) every 169 (122-213) min, the absolute relative deviation was lower (7.0% [3.5, 13.0] vs. 12.8% [6.3, 21.8], P<0.001), and the percentage of points in the Clarke error grid Zone A was higher (87.8% vs. 70.2%).

CONCLUSIONS:

Accuracy of the Navigator CGM device during critical illness was comparable to that observed in non-critical care settings. Further significant improvements in accuracy may be obtained by frequent calibrations with ABG measurements.

PMID:
24180327
PMCID:
PMC3894676
DOI:
10.1089/dia.2013.0221
[Indexed for MEDLINE]