Clin Nutr. 2014 Oct;33(5):867-71. doi: 10.1016/j.clnu.2013.10.007. Epub 2013 Oct 23.

Performance assessment of a glucose control protocol in septic patients with an automated intermittent plasma glucose monitoring device.

Author information

1
Unità Operativa di Anestesia e Rianimazione, Azienda Ospedaliera San Paolo - Polo Universitario, Italy. Electronic address: michele.umbrello@ao-sanpaolo.it.
2
Dipartimento di Fisiopatologia Medico-Chirurgica e dei Trapianti, Università degli Studi di Milano, Italy.
3
Unità Operativa di Anestesia e Rianimazione, Azienda Ospedaliera San Paolo - Polo Universitario, Italy.
4
Dipartimento di Scienze della Salute, Università degli Studi di Milano, Italy.
5
Unità Operativa di Anestesia e Rianimazione, Azienda Ospedaliera San Paolo - Polo Universitario, Italy; Dipartimento di Fisiopatologia Medico-Chirurgica e dei Trapianti, Università degli Studi di Milano, Italy.

Abstract

BACKGROUND & AIMS:

The optimal level and modality of glucose control in critically ill patients is still debated. A protocolized approach and the use of nearly-continuous technologies are recommended to manage hyperglycemia, hypoglycemia and glycemic variability. We recently proposed a pato-physiology-based glucose control protocol which takes into account patient glucose/carbohydrate intake and insulin resistance. Aim of the present investigation was to assess the performance of our protocol with an automated intermittent plasma glucose monitoring device (OptiScanner™ 5000).

METHODS:

OptiScanner™ was used in 6 septic patients, providing glucose measurement every 15' from a side-port of an indwelling central venous catheter. Target level of glucose was 80-150 mg/dL. Insulin infusion and kcal with nutritional support were also recorded.

RESULTS:

6 septic patients were studied for 319 h (1277 measurements); 58 [45-65] hours for each patient (measurements/patient: 231 [172-265]). Blood glucose was at target for 93 [90-98]% of study time. Mean plasma glucose was 126 ± 11 mg/dL. Only 3 hypoglycemic episodes (78, 78, 69 mg/dL) were recorded. Glucose variability was limited: plasma glucose coefficient of variation was 11.7 ± 4.0% and plasma glucose standard deviation was 14.3 ± 5.5 mg/dL.

CONCLUSIONS:

The local glucose control protocol achieved satisfactory glucose control in septic patients along with a high degree of safeness. Automated intermittent plasma glucose monitoring seemed useful to assess the performance of the protocol.

KEYWORDS:

Automated glucose control; Blood glucose; Critically ill patients; Glucose control protocol; Sepsis