JACC Cardiovasc Interv. 2013 Sep;6(9):955-64. doi: 10.1016/j.jcin.2013.05.008. Epub 2013 Aug 14.

Automated 3-dimensional aortic annular assessment by multidetector computed tomography in transcatheter aortic valve implantation.

Author information

1
Institut Cardiovasculaire Paris, Hôpital Privé Jacques Cartier, Générale de Santé, Massy, France.

Abstract

OBJECTIVES:

This study sought to evaluate the accuracy, reproducibility, and predictive value for post-procedural aortic regurgitation (AR) of an automated multidetector computed tomography (MDCT) post-processing imaging software, 3mensio Valves (version 5.1.sp1, 3mensio Medical Imaging BV, the Netherlands), in the assessment of patients undergoing transcatheter aortic valve implantation (TAVI).

BACKGROUND:

Accurate pre-operative aortic annulus measurements are crucial for patients undergoing TAVI.

METHODS:

One hundred five patients undergoing MDCT screening before TAVI were evaluated. Aortic annular measurement was compared between automated 3mensio Valves software and manual data post-processing software on a dedicated workstation; we analyzed the discrimination value of annulus measurement for post-procedural AR in 44 recipients of a self-expanding valve.

RESULTS:

The automated 3mensio Valves software showed good concordance with manual MDCT measurements as demonstrated by Bland-Altman analysis. The automated software provided equally good reproducibility as manual measurement, especially for measurement of aortic annulus area (intraobserver intraclass correlation coefficients 0.98 vs. 0.97, interobserver 0.98 vs. 0.95). In 44 patients after implantation of a self-expanding valve, the valve diameter/CT-measured geometric mean annulus diameter ratio by automated 3mensio Valves software showed moderate and better discrimination ability in predicting post-procedural AR compared with manual measurement (p = 0.12, area under the curve 0.77, 95% confidence interval: 0.63 to 0.91, area under the curve 0.68, 95% confidence interval: 0.50 to 0.86, respectively).

CONCLUSIONS:

The automated 3mensio Valves software demonstrated reliable, reproducible aortic annulus measurement and better predictive value for post-procedural AR, suggesting important clinical implications for pre-operative assessment of patients undergoing TAVI.

KEYWORDS:

AR; AS; AUC; CI; ICC; MDCT; ROC; TAVI; TEE; annulus measurement; aortic regurgitation; aortic stenosis; area under the curve; confidence interval; geometric mean annulus diameter measured by computed tomography; intraclass correlation coefficient; lDiam-CT; long-axis annulus diameter measured by computed tomography; mDiam-CT; multidetector computed tomography; receiver-operating characteristic; sDiam-CT; short-axis annulus diameter measured by computed tomography; transcatheter aortic valve implantation; transesophageal echocardiography

PMID:
23954060
DOI:
10.1016/j.jcin.2013.05.008
[Indexed for MEDLINE]