Scanning ion sensitive probe for plasma profile measurements in the boundary of the Alcator C-Mod tokamak

Rev Sci Instrum. 2013 May;84(5):053507. doi: 10.1063/1.4807699.

Abstract

A new Ion Sensitive Probe head has been created for the outer-midplane scanning probe system on the Alcator C-Mod tokamak. The new probe head contains three elements: an ion sensitive probe to measure ion temperature and plasma potential, a Langmuir probe to measure electron temperature, density, and floating potential, and a second Langmuir probe to measure ion saturation current and the density fluctuations arising from ''blob'' events. The ion sensitive probe current is normalized to this measurement to reduced deleterious effects of the strong fluctuations. Design of the high heat flux probe (>100 MW/m(2)) and initial results are presented.