Appl Radiat Isot. 2013 Nov;81:242-7. doi: 10.1016/j.apradiso.2013.03.093. Epub 2013 Apr 2.

Determination of blank indication of active radon monitors.

Author information

  • 1ENEA, Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti (INMRI), Centro Ricerche Casaccia, I-00123 Rome, Italy. francesco.cardellini@enea.it

Abstract

The importance of determination of the blank indication for active radon monitors is discussed. Two new blank chambers were developed at ENEA-INMRI that allows such determination with an uncertainty lower than a few Bq/m(3). Results are reported and that show the large blank variability between radon active monitors and its growth with time when such monitors are used at high radon concentrations.

Copyright © 2013 Elsevier Ltd. All rights reserved.

KEYWORDS:

(210)Pb poisoning; Active radon monitor; Blank indication; Radon chamber; Radon detector background; Radon detector contamination