Food Chem. 2013 Aug 15;139(1-4):1029-35. doi: 10.1016/j.foodchem.2013.01.071. Epub 2013 Feb 8.

A fast, reliable, ultra high performance liquid chromatography method for the simultaneous determination of amino acids, biogenic amines and ammonium ions in cheese, using diethyl ethoxymethylenemalonate as a derivatising agent.

Author information

1
Instituto de Productos Lácteos de Asturias (IPLA-CSIC), 33300-Villaviciosa, Asturias, Spain. bredruel@ipla.csic.es

Abstract

Derivatisation treatment with diethyl ethoxymethylenemalonate followed by ultra-HPLC allowed the simultaneous quantification of 22 amino acids, 7 biogenic amines and ammonium ions in cheese samples in under 10 min. This is the fastest elution time ever reported for such a resolution. The proposed method shows good linearity (R(2)>0.995) and sensitivity (detection limit 0.08-3.91 μM; quantification limit <13.02 μM). Intra- and inter-day repeatability ranged from 0.35% to 1.25% and from 0.85% to 5.2%, respectively. No significant effect of the cheese matrix was observed.

PMID:
23561206
DOI:
10.1016/j.foodchem.2013.01.071
[Indexed for MEDLINE]