Magn Reson Med. 2014 Jan;71(1):95-104. doi: 10.1002/mrm.24633. Epub 2013 Mar 15.

High-resolution 3D radial bSSFP with IDEAL.

Author information

1
Department of Medical Physics, University of Wisconsin, Madison, Wisconsin, USA.

Abstract

Radial trajectories facilitate high-resolution balanced steady state free precession (bSSFP) because the efficient gradients provide more time to extend the trajectory in k-space. A number of radial bSSFP methods that support fat-water separation have been developed; however, most of these methods require an environment with limited B0 inhomogeneity. In this work, high-resolution bSSFP with fat-water separation is achieved in more challenging B0 environments by combining a 3D radial trajectory with the IDEAL chemical species separation method. A method to maintain very high resolution within the timing constraints of bSSFP and IDEAL is described using a dual-pass pulse sequence. The sampling of a unique set of radial lines at each echo time is investigated as a means to circumvent the longer scan time that IDEAL incurs as a multiecho acquisition. The manifestation of undersampling artifacts in this trajectory and their effect on chemical species separation are investigated in comparison to the case in which each echo samples the same set of radial lines. This new bSSFP method achieves 0.63 mm isotropic resolution in a 5-min scan and is demonstrated in difficult in vivo imaging environments, including the breast and a knee with ACL reconstruction hardware at 1.5 T.

KEYWORDS:

IDEAL; balanced steady state free precession; radial imaging; water-fat imaging

PMID:
23504943
PMCID:
PMC3762898
DOI:
10.1002/mrm.24633
[Indexed for MEDLINE]
Free PMC Article
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for Wiley Icon for PubMed Central
    Loading ...
    Support Center