Format

Send to

Choose Destination
See comment in PubMed Commons below
J Hosp Infect. 2013 Apr;83(4):344-6. doi: 10.1016/j.jhin.2013.01.003. Epub 2013 Feb 16.

Outbreak of multidrug-resistant Pseudomonas aeruginosa infection following urodynamic studies traced to contaminated transducer.

Author information

1
Division of Epidemiology and Preventive Medicine, Tel-Aviv Sourasky Medical Center, Tel Aviv, Israel. yoji@netvision.net.il

Abstract

We report a nosocomial outbreak of urinary tract infection by extremely drug resistant Pseudomonas aeruginosa, susceptible only to colistin. Infection in three patients followed urodynamic studies. Two of the three patients were children, one of whom also developed urosepsis. The investigation led to detection of contaminated pressure transducers. Genotyping confirmed that patient and transducer isolates were identical. These transducers were not labelled as 'single use only' despite the possibility that contaminated urine may reflux and mix with the fluid in the device. The issue of re-usable versus single-use urodynamic devices is discussed.

PMID:
23419597
DOI:
10.1016/j.jhin.2013.01.003
[Indexed for MEDLINE]
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for Elsevier Science
    Loading ...
    Support Center