Format

Send to

Choose Destination
Conf Proc IEEE Eng Med Biol Soc. 2012;2012:5642-5. doi: 10.1109/EMBC.2012.6347274.

Heart rate variability analysis using a seismocardiogram signal.

Author information

1
Group of Biomedical and Electronic Instrumentation of the Department of Electronic Engineering of the Universitat Politècnica de Catalunya (UPC), Barcelona, 08034 Spain. jramos@eel.upc.edu

Abstract

Seismocardiography is a simple and non invasive method of recording cardiac activity from the movements of the body caused by heart pumping. In this preliminary study we use a smartphone to record this acceleration and estimate the heart rate. We compare the heart rate variability parameters from the seismocardiogram and ECG reference signal. The results show a great similarity and are strongly influenced by the instability in the sampling frequency of the device. The differences between RR series are lower than 10 ms.

PMID:
23367209
DOI:
10.1109/EMBC.2012.6347274