Format

Send to

Choose Destination
See comment in PubMed Commons below
Appl Opt. 2012 Oct 1;51(28):6855-63. doi: 10.1364/AO.51.006855.

Design of a flat field concave-grating-based micro-Raman spectrometer for environmental applications.

Author information

1
School of Biomedical Engineering, McMaster University, 1280 Main Street West, Hamilton, Ontario L8S 4K1, Canada.

Abstract

In order to simplify the design process of microfabricated concave gratings, simplified algorithms for fast characterization of the concave grating were developed. These algorithms can be used to assist system designers using ray-tracing software in the determination of optimum design parameters considering the requirements and restrictions for specific applications. According to the algorithms, it is feasible to design a flat field microconcave grating with a 4 mm grating radius as a key component in a micro-Raman spectrometer system for inline environmental monitoring applications. This microspectrometer operates over the spectral wavelength band from 785 nm to 1000 nm and has a spectral resolution of 2 nm at 900 nm. The total size of the system is 1 mm×4 mm×3.7 mm, making it one of the smallest for this wavelength range and spectrum resolution.

PMID:
23033103
[Indexed for MEDLINE]
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Loading ...