ZSI 375 artificial urinary sphincter for male urinary incontinence: a preliminary study

BJU Int. 2013 Apr;111(4 Pt B):E202-6. doi: 10.1111/j.1464-410X.2012.11468.x. Epub 2012 Sep 3.

Abstract

Objective: To assess retrospectively the safety and efficacy of an artificial urinary sphincter, the ZSI 375 device (Zephyr Surgical Implants, Geneva, Switzerland), in male patients with moderate-to-severe stress urinary incontinence after a prostate or bladder intervention.

Patients and methods: The ZSI 375 device is a one-piece device consisting of an adjustable cuff, moulded to fit around the urethra, which is connected by a tube to a pump and a pressure-regulating tank. It has no abdominal reservoir. Patients underwent a perineal incision for cuff placement and an inguinal incision for pump and tank scrotal placement. Complications and pads used to manage incontinence were recorded.

Results: Between May 2009 and April 2011, 36 patients underwent ZSI 375 device placement. The median (range) follow-up was 15.4 (6-28) months. No patient experienced bladder overactivity, chronic urinary retention, or any other adverse effect after device activation. Complications leading to device removal arose in four patients (one case of erosion, three cases of infection). Social continence (0 or 1 pad/day) was achieved in 28/36 patients (78%) at 3 months and 26/36 patients (73%) at 6 months after device activation. In 12/14 patients for a sphincter closure pressure range of 60-70 cm H2O, in 3/3 patients for a range of 70-80 cm H2O and in 2/11 for a range of 90-100 cm, H2O social continence was achieved only after increasing the pressure of the cuff by trans-scrotal injection of saline.

Conclusions: The ZSI 375 device is safe and effective but our follow-up may not have been long enough to identify all potential complications. Further research is needed to confirm these results and extend our investigation, for instance, to the peno-scrotal approach.

Publication types

  • Comparative Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Follow-Up Studies
  • Humans
  • Male
  • Middle Aged
  • Prosthesis Design
  • Prosthesis Implantation / methods*
  • Retrospective Studies
  • Treatment Outcome
  • Urinary Incontinence / physiopathology
  • Urinary Incontinence / surgery*
  • Urinary Sphincter, Artificial*