FLUKA Monte Carlo simulation for the Leksell Gamma Knife Perfexion radiosurgery system: homogeneous media

Phys Med. 2013 Nov;29(6):656-61. doi: 10.1016/j.ejmp.2012.07.005. Epub 2012 Aug 14.

Abstract

The purpose of this work is to investigate the capability of the FLUKA Monte Carlo (MC) code to simulate the Elekta Leksell Gamma Knife Perfexion (LGK-PFX) and reproduce the Treatment Planning System (TPS) Leksell GammaPlan version 8.2 (LGP) dose calculations for the case of a water equivalent phantom target. Thanks to the collaboration with Elekta Instruments AB, the collimation system geometry, the source positions and all the involved material have been simulated in detail. The relative linear dose distribution along the three coordinate axes, for each collimator size, and the Relative Output Factors (ROF) have been investigated. The simulation has been validated comparing simulated linear dose profiles with measurements performed with EBT radiochromic films. The acceptance criterion between experimental data and FLUKA results is based on the gamma index (GI) method. The FLUKA MC calculation for the ROF provided the values of 0.920 for the 8 mm collimators and 0.800 for the 4 mm collimators. These values are in good agreement with the Elekta reference data of 0.924 and 0.805 respectively. The percentage difference between calculated and reference values for the ROF is under 1% and within the FLUKA uncertainty. Also the simulated relative dose profiles show a good agreement with the LGP calculation expressed by means of the gamma index method. This established accuracy proves that FLUKA is a suitable and powerful tool in order to reproduce successfully the LGP calculations for the homogeneous media.

Keywords: FLUKA; Gamma Knife Perfexion; Monte Carlo; Stereotactic radiosurgery.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Monte Carlo Method*
  • Radiation Dosage
  • Radiosurgery*