Format

Send to

Choose Destination
J Clin Microbiol. 2012 Sep;50(9):3105-7. doi: 10.1128/JCM.00730-12. Epub 2012 Jul 11.

Investigation of false-positive results given by the QuantiFERON-TB Gold In-Tube assay.

Author information

1
Division of Infectious Diseases and Geographic Medicine, Department of Medicine, Stanford University School of Medicine, Palo Alto, California, USA.

Abstract

We investigated a sudden increase in the rate of positive QuantiFERON-TB Gold In-Tube results from 10% to 31% at a U.S. academic institution. Direct comparison of the TB antigen tubes with tubes from a different lot number identified that a potential problem with the TB antigen vials in a certain tube lot was the likely cause of the elevated positive rate. The underlying defect remains unknown. This finding warrants refinement of quality control programs by the manufacturer and users.

PMID:
22785197
PMCID:
PMC3421800
DOI:
10.1128/JCM.00730-12
[Indexed for MEDLINE]
Free PMC Article