Abstract

OBJECTIVES:

Transapical (TA) aortic valve implantation (AVI) has evolved as an alternative procedure for high-risk patients. We evaluated the second-generation SAPIEN XT™ prosthesis in a prospective multicentre clinical trial.

METHODS:

A total of 150 patients (age: 81.6 ± 5.8 years; 40.7% female) were included. Prosthetic valves (diameter: 23 mm (n = 36), 26 mm (n = 57) and 29 mm (n = 57)) were implanted. The ASCENDRA-II™ modified delivery system was used in the smaller sizes. Mean logistic EuroSCORE was 24.3 ± 7.0%, and mean STS score 7.5 ± 4.4%. All patients gave written informed consent.

RESULTS:

Off-pump AVI was performed using femoral arterial and venous access wires as a safety net. All but two patients received TA-AVI, as planned. The 29-mm valve showed similar function as the values of two other diameters did. Three patients (2%) required temporary cardiopulmonary bypass support. Postoperative complications included renal failure requiring long-term dialysis in four, bleeding requiring rethoracotomy in four, respiratory complication requiring reintubation in eight and sepsis in four patients, respectively. Thirty-day mortality was 13 (8.7%) for the total cohort and 2/57 (3.5%) for patients receiving the 29-mm valve, respectively. Echocardiography at discharge showed none or trivial aortic incompetence (AI) in 71% and mild-AI in 22% of the patients. Post-implantation AI was predominantly paravalvular and ≥ 2+ in 7% of patients. One patient required reoperation for AI within 30 days.

CONCLUSIONS:

The PREVAIL TA multicentre trial demonstrates good functionality and good outcomes for TA-AVI, using the SAPIEN XT™ prosthesis and its second-generation ASCENDRA-II™ delivery system, as well successful introduction of the 29-mm SAPIEN XT™ valve for the benefit of high-risk elderly patients.