Toxicol Mech Methods. 2012 Apr;22(3):205-10. doi: 10.3109/15376516.2011.610386. Epub 2011 Sep 30.

An analytical procedure for the determination of aluminum used in antiperspirants on human skin in Franz™ diffusion cell.

Author information

1
CHU Poitiers, Laboratoire de Biochimie, Poitiers, France. olivier.guillard@univ-poitiers.fr

Abstract

A local case report of hyperaluminemia (aluminum concentration: 3.88 µmol/L) in a woman using an aluminum-containing antiperspirant for 4 years raises the question of possible transdermal uptake of aluminum salt as a future public health problem. Prior to studying the transdermal uptake of three commercialized cosmetic formulas, an analytical assay of aluminum (Al) in chlorohydrate form (ACH) by Zeeman Electrothermal Atomic Absorption Spectrophotometer (ZEAAS) in a clean room was optimized and validated. This analysis was performed with different media on human skin using a Franz(™) diffusion cell. The detection and quantification limits were set at ≤ 3 µg/L. Precision analysis as within-run (n = 12) and between-run (n = 15-68 days) yield CV ≤ 6%. The high analytic sensitivity (2-3 µg/L) and low variability should allow an in vitro study of the transdermal uptake of ACH.

PMID:
21962178
DOI:
10.3109/15376516.2011.610386
[Indexed for MEDLINE]