Head Neck. 2011 Sep;33(9):1285-93. doi: 10.1002/hed.21597. Epub 2010 Oct 27.

Endotracheal temperature and humidity in laryngectomized patients in a warm and dry environment and the effect of a heat and moisture exchanger.

Author information

1
Department of Head and Neck Oncology and Surgery, The Netherlands Cancer Institute-Antoni van Leeuwenhoek Hospital, Amsterdam, The Netherlands.

Abstract

BACKGROUND:

This study was designed to assess endotracheal climate in laryngectomized patients in a warm and dry environment and the effects of a heat and moisture exchanger (HME).

METHODS:

Endotracheal temperature and humidity were measured in 11 laryngectomized patients with a regularly used HME (Provox Normal HME; R-HME), an HME with antimicrobial filter (Provox Micron HME; F-HME), and without HME (open stoma). Measurements were performed at 32°C, 34°C, and 38°C (<25% relative humidity).

RESULTS:

Both R-HME and F-HME increased end-inspiratory humidity (AH(insp)) equally (range, 3.4 to 5.6 mg H(2)O/L). The R-HME has a cooling effect on end-inspiratory temperature (T(insp)), which is similar for all tested environmental conditions (-3.8°C); F-HME decreases T(insp) less (range, -1.3 to -0.6°C).

CONCLUSIONS:

In a warm and dry environment, both R- and F-HME significantly cool and humidify inspired air. Therefore, consistent use of an HME under these climate conditions is also probably clinically beneficial.

PMID:
21837698
DOI:
10.1002/hed.21597
[Indexed for MEDLINE]