Dose reduction using bismuth shielding during paediatric CT examinations in Slovakia

Radiat Prot Dosimetry. 2011 Sep;147(1-2):160-3. doi: 10.1093/rpd/ncr336. Epub 2011 Aug 4.

Abstract

Considering the massive increase of computer tomography (CT) examinations in Slovakia during the last 10 y, it can be expected that a higher radiation load may be observed in the Slovak population. Since child population is more sensitive to radiation than adult population, a monitoring has started to see how high the radiation dose is for paediatric patients during CT examinations in chosen departments in Slovakia. The CT examination of the head is one of the most frequently done examinations in Slovakian departments and that is why measurements were done to clarify how usage of bismuth shields for eyes and thyroid can affect the eye and thyroid doses. For simulation, 215 thermoluminescent dosimeters were exposed on anthropomorphic phantom of a child with and without usage of bismuth shields. The result was that only two of the three chosen departments confirmed a reduction. On the other hand, one of the departments confirmed that the reduction can be up to 56-65 %, which is significant.