J Magn Reson Imaging. 2011 Sep;34(3):702-9. doi: 10.1002/jmri.22664. Epub 2011 Jul 18.

Motion-sensitized driven equilibrium for blood-suppressed T2* mapping.

Author information

1
GSK Clinical Imaging Centre, Hammersmith Hospital, London, United Kingdom. rexford.newbould@gmail.com

Abstract

PURPOSE:

To combine a motion-sensitized driven equilibrium (MSDE) preparation with a multi-echo spoiled gradient-echo sequence (SPGR) to suppress the blood signal intensity in T2* mapping of carotid plaques and liver.

MATERIALS AND METHODS:

The analytical solution of the Bloch equations for the multi-echo SPGR sequence, with and without the MSDE preparation, was calculated to optimize sequence parameters. The sequence was implemented at 3 Tesla and T2* maps of carotid plaques and liver were first optimized for blood suppression, and then acquired from five healthy livers and six subjects with ultrasound-identified carotid plaques.

RESULTS:

Simulations and experimental data showed that the flip angle that gives maximal signal (Ernst angle) in the MSDE-SPGR was greatly increased from that of pure SPGR. Robust blood suppression was obtained in the T2* maps of carotid plaques and liver, requiring suppression at a field of speed (FOS) of 30 cm/s in both the carotids and liver.

CONCLUSION:

MSDE provides a means to suppress the blood signal intensity in SPGR sequences. Tissue T2* maps can be obtained without the confounding effects originating from blood vessels.

PMID:
21769964
DOI:
10.1002/jmri.22664
[Indexed for MEDLINE]