Narrow-bore core-shell particles and monolithic columns in the analysis of silybin diastereoisomers

J Sep Sci. 2011 Aug;34(16-17):2206-13. doi: 10.1002/jssc.201100309. Epub 2011 Jun 27.

Abstract

Two chromatographic narrow-bore columns, a novel 2.6 μm particle-packed Kinetex™ C18 core-shell (50×2.1 mm id) and monolithic Chromolith(®) FastGradient RP-18e (50×2 mm id), were evaluated for the analysis of diastereoisomers of the flavonolignans silybin and 23-O-acetylsilybin under isocratic conditions. The main advantages of the core-shell column are markedly higher efficiency (hmin =2.8 versus 5.6 for silybin A) and better peak symmetry. The Kinetex column exhibits only a slight change in the height equivalent of the theoretical plate with a higher linear velocity of the mobile phase. The monolithic column shows notably higher selectivity in terms of selectivity factor (1.21 versus 1.12) in the analysis of critical-pair of diastereoisomers (silybin A and silybin B) and enables shorter run duration (approx. twofold) together with lower backpressure. The resolution power was found to be comparable, but the Kinetex column required a higher pressure of the mobile phase that, together with the higher chance of clogging, can be a disadvantage in the separation of biological samples. Successful baseline separation of silybin diastereoisomers in real pharmaceutical sample on monolithic column was accomplished.

Keywords: Chromolith; Core-shell; Kinetex; Monolithic column; Silybin.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Chromatography, High Pressure Liquid / instrumentation
  • Chromatography, High Pressure Liquid / methods*
  • Resins, Synthetic / chemistry*
  • Silybin
  • Silymarin / chemistry*
  • Silymarin / isolation & purification
  • Stereoisomerism

Substances

  • Resins, Synthetic
  • Silymarin
  • Silybin