Format

Send to

Choose Destination
J Clin Microbiol. 2011 Aug;49(8):3054-7. doi: 10.1128/JCM.01035-11. Epub 2011 Jun 8.

Controlled comparison of BacT/Alert MB system, manual Myco/F lytic procedure, and isolator 10 system for diagnosis of Mycobacterium tuberculosis Bacteremia.

Author information

1
Box 102359, Duke University Medical Center, Durham, NC 27710, USA. john.crump@duke.edu

Abstract

We compared the performance of the BacT/Alert MB system, that of the manual Bactec Myco/F Lytic procedure, and that of the Isolator 10 lysis-centrifugation system in the detection of Mycobacterium tuberculosis bacteremia. Mean times to detection were 16.4 days for BacT/Alert MB versus 20.0 days for Myco/F Lytic, 16.5 days for BacT/Alert MB versus 23.8 days for Isolator 10, and 21.1 days for Bactec Myco/F Lytic versus 22.7 days for Isolator 10. There were no significant differences in yields. The mean (range) magnitude of mycobacteremia was 30.0 (0.4, 90.0) CFU/ml and was correlated with the time to positivity in the BacT/Alert MB system (r = -0.4920). M. tuberculosis bacteremia was detected more rapidly in a continuously monitored liquid blood culture system, but the mean time to positivity exceeded 3 weeks.

PMID:
21653761
PMCID:
PMC3147725
DOI:
10.1128/JCM.01035-11
[Indexed for MEDLINE]