Format

Send to

Choose Destination
Anal Chem. 2011 Jul 1;83(13):5442-6. doi: 10.1021/ac200812d. Epub 2011 Jun 14.

Performance characteristics of a new hybrid quadrupole time-of-flight tandem mass spectrometer (TripleTOF 5600).

Author information

1
W. M. Keck FT-ICR Mass Spectrometry Laboratory, Department of Chemistry, North Carolina State University, Raleigh, North Carolina 27695, USA.

Abstract

The TripleTOF 5600 System, a hybrid quadrupole time-of-flight mass spectrometer, was evaluated to explore the key figures of merit in generating peptide and protein identifications that included spectral acquisition rates, data quality, proteome coverage, and biological depth. Employing a Saccharomyces cerevisiae tryptic digest, careful consideration of several performance features demonstrated that the speed of the TripleTOF contributed most to the resultant data. The TripleTOF system was operated with 8, 20, and 50 MS/MS events in an effort to compare with other MS technologies and to demonstrate the abilities of the instrument platform.

PMID:
21619048
PMCID:
PMC3138073
DOI:
10.1021/ac200812d
[Indexed for MEDLINE]
Free PMC Article

Supplemental Content

Full text links

Icon for American Chemical Society Icon for PubMed Central
Loading ...