JACC Cardiovasc Interv. 2011 Apr;4(4):452-9. doi: 10.1016/j.jcin.2010.11.016.

Randomized comparison of a polymer-free sirolimus-eluting stent versus a polymer-based paclitaxel-eluting stent in patients with diabetes mellitus: the LIPSIA Yukon trial.

Author information

1
Department of Internal Medicine/Cardiology, University of Leipzig, Heart Center, Leipzig, Germany. stdesch@web.de

Abstract

OBJECTIVES:

The objective of the study was to assess noninferiority of the polymer-free sirolimus-eluting Yukon Choice stent (Translumina GmbH, Hechingen, Germany) compared with the polymer-based Taxus Liberté stent (Boston Scientific, Natick, Massachusetts) with regard to the primary endpoint, in-stent late lumen loss, at 9 months in patients with diabetes mellitus.

BACKGROUND:

The Yukon Choice stent has been evaluated in several randomized controlled trials before, albeit to date, there has been no trial that exclusively enrolled patients with diabetes mellitus.

METHODS:

Patients with diabetes mellitus undergoing percutaneous coronary intervention for clinically significant de novo coronary artery stenosis were randomized 1:1 to receive either the polymer-free sirolimus-eluting Yukon Choice stent or the polymer-based paclitaxel-eluting Taxus Liberté stent.

RESULTS:

A total of 240 patients were randomized. Quantitative coronary angiography was available for 79% of patients. Mean in-stent late lumen loss was 0.63 ± 0.62 mm for the Yukon Choice stent and 0.45 ± 0.60 mm for the Taxus Liberté stent. Based on the pre-specified margin, the Yukon Choice stent failed to show noninferiority for the primary endpoint. During follow-up, there were no significant differences between groups regarding death, myocardial infarction, stent thrombosis, target lesion revascularization, target vessel revascularization, or nontarget vessel revascularization.

CONCLUSIONS:

Compared with the Taxus Liberté stent, the polymer-free sirolimus-eluting Yukon Choice stent failed to show noninferiority with regard to the primary endpoint, in-stent late lumen loss, in patients with diabetes mellitus after 9-month follow-up. Both stents showed comparable clinical efficacy and safety. (Yukon Choice Versus Taxus Liberté in Diabetes Mellitus; NCT00368953).