FBX aqueous chemical dosimeter for measurement of dosimetric parameters

Appl Radiat Isot. 2011 Feb;69(2):399-402. doi: 10.1016/j.apradiso.2010.10.001. Epub 2010 Oct 8.

Abstract

We investigated the ferrous sulphate-benzoic acid-xylenol orange (FBX) aqueous chemical dosimeter for measurement of dosimetric parameters such as the output factor, backscatter factor and lateral beam profiles for different square fields sizes for (60)Co γ-rays. A water phantom was employed to measure these parameters. An ionization chamber (IC) was used for calibration and comparison. A comparison of the resulting measurements with an ionization chamber's measured parameters showed good agreement. We thus believe that the tissue equivalent FBX dosimetry system can measure the dosimetric parameters for (60)Co with reasonable accuracy.

MeSH terms

  • Benzoic Acid
  • Cobalt Radioisotopes
  • Ferrous Compounds
  • Gamma Rays
  • Phantoms, Imaging
  • Phenols
  • Radiometry / methods*
  • Sulfoxides
  • Water
  • Xylenes

Substances

  • Cobalt Radioisotopes
  • Ferrous Compounds
  • Phenols
  • Sulfoxides
  • Xylenes
  • Water
  • ferrous sulfate
  • Benzoic Acid
  • xylenol orange