Appl Opt. 1994 Dec 1;33(34):7823-8. doi: 10.1364/AO.33.007823.

Performance of a circle-to-line optical system for a Fabry-Perot interferometer: a laboratory study.

Abstract

We describe the characterization and laboratory study of a new optical device, a circle-to-line interferometer optical (CLIO) system, for a Fabry-Perot interferometer. The CLIO system converts a circular Fabry-Perot interferometer fringe pattern into a linear fringe pattern that can be analyzed by a linear array detector or a charge-coupled device. One can achieve the circle-to-line conversion by the use of a mirrored kaleidoscope and a 90-deg segment of a 45-deg half-angle internally reflecting cone. Our laboratory results are in good agreement with theoretical predictions. The aberrations associated with this optical system are minor for large F-number optical systems, especially for the cone segment. The finesse degradation caused by this CLIO system is relatively small, which agrees with the results of a ray-trace study.