J Glaucoma. 2011 Oct;20(8):470-6. doi: 10.1097/IJG.0b013e3181f3eb64.

Reproducibility of retinal nerve fiber layer thickness measurements using spectral domain optical coherence tomography.

Author information

  • 1Glaucoma Service, Massachusetts Eye and Ear Infirmary, Department of Ophthalmology, Harvard Medical School, Boston, MA 02114, USA.

Abstract

PURPOSE:

To evaluate the reproducibility of the peripapillary retinal nerve fiber layer (RNFL) thickness measurements obtained by Spectralis spectral domain optical coherence tomography (OCT) (Heidelberg Engineering, Heidelberg, Germany) in normal and glaucoma participants.

METHODS:

Participants were recruited from a university-based clinic. Peripapillary RNFL thickness measurements were repeated 3 times during the same visit using the follow-up function. One eye of each participant was randomly selected for statistical analysis. Reproducibility was evaluated using within-subject standard deviation (Sw), coefficient of variation (CV), and intraclass correlation coefficient (ICC). Spearman rank correlation coefficient analyses were used to assess the correlation of the standard deviation of the 3 measurements for each participant with the RNFL thickness value.

RESULTS:

Forty-five normal participants and 33 glaucoma patients were included in the study. The CVs ranged from 1.45% [overall global (G)] to 2.59% [temporal quadrant (T)] in normal eyes and from 1.74% (G) to 3.22% (T) in the glaucomatous eyes. ICCs ranged from 0.977 (T) to 0.990 (G and inferior-nasal sector) in normal eyes and from 0.983 (T) to 0.997 (inferior quadrant) in glaucomatous eyes. Sw were from 1.34 μm (G) to 2.39 μm (superior-temporal and inferior-temporal sectors) in normal eyes and from 1.14 μm (G) to 2.25 μm (superior-nasal sector) in the glaucomatous eyes. There were no significant correlations between RNFL thickness values and the measurement variability for each participant.

CONCLUSIONS:

Spectralis OCT shows excellent reproducibility for measuring the peripapillary RNFL thickness values in both healthy and glaucoma participants.

PMID:
20852437
PMCID:
PMC3500562
DOI:
10.1097/IJG.0b013e3181f3eb64
[PubMed - indexed for MEDLINE]
Free PMC Article
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for Lippincott Williams & Wilkins Icon for PubMed Central
    Loading ...
    Support Center