Interact Cardiovasc Thorac Surg. 2010 Dec;11(6):740-4. doi: 10.1510/icvts.2010.242123. Epub 2010 Sep 19.

Early diagnosis of lung cancer using a SAFE-3000 autofluorescence bronchoscopy.

Author information

1
Department of Thoracic Surgery, University of L'Aquila, G. Mazzini Hospital, Piazza Italia 1, Circonvallazione Ragusa 39, 64100 Teramo, Italy. duilio.divisi@aslteramo.it

Abstract

The aim of the study was to evaluate the double modality working of a new autofluorescence videobronchoscope SAFE-3000 for the detection and localization of precancerous and malignant lesions. From April to May 2009, 168 patients underwent SAFE-3000 bronchoscopy using 'Twin Mode' and 'Multiple Image Xposition (MIX)' technologies. The study considers only 97 patients with morphological alterations (visual score 2 or 3) of mucosa; four bronchial biopsies (two for every modality) have been performed on every patient. Histological diagnosis of the 388 biopsies revealed normal mucosa in 11 patients, four benign lesions, 27 precancerous lesions and 55 malignant lesions. The sensitivity of the 'Twin Mode' and 'MIX' techniques in the characterization of premalignant and malignant lesions was 96% vs. 100%. The specificity was 60% in both of these technologies. SAFE-3000 autofluorescence bronchoscopy allows an early diagnosis of preneoplastic or neoplastic lesions according to the careful analysis of the bronchial mucosa, due to the complementarity of two modalities 'Twin Mode' and 'MIX'. Based on the relationship between the bronchoscopic goal and the histological results the MIX method contributes more of the new endoscopic technique findings.

PMID:
20852332
DOI:
10.1510/icvts.2010.242123
[Indexed for MEDLINE]

Supplemental Content

Full text links

Icon for Silverchair Information Systems
Loading ...
Support Center