Optimized EPI for fMRI using a slice-dependent template-based gradient compensation method to recover local susceptibility-induced signal loss

MAGMA. 2010 Jun;23(3):165-76. doi: 10.1007/s10334-010-0215-x. Epub 2010 May 21.

Abstract

Object: Most functional magnetic resonance imaging (fMRI) experiments use gradient-echo echo planar imaging (GE EPI) to detect the blood oxygenation level-dependent (BOLD) effect. This technique may fail in the presence of anatomy-related susceptibility-induced field gradients in the human head. In this work, we present a novel 3D compensation method in combination with a template-based correction that can be optimized over particular regions of interest to recover susceptibility-induced signal loss without acquisition time penalty.

Materials and methods: Based on an evaluation of B(0) field maps of eight subjects, slice-dependent gradient compensation moments are derived for maximal BOLD sensitivity in two compromised regions: the orbitofrontal cortex and the amygdala areas. A modified EPI sequence uses these additional gradient moments in all three imaging directions. The method is compared to non-compensated, template-based and subject-specific correction gradients and also in a breath-holding experiment.

Results: The slice-dependent gradient compensation method significantly improves signal intensity/BOLD sensitivity by about 35/43% in the orbitofrontal cortex and by 17/30% in the amygdala areas compared to a conventional acquisition. Template-based correction and subject-specific correction perform equally well. The BOLD sensitivity in the breath hold experiment is effectively increased in compensated regions.

Conclusion: The new method addresses the problem of susceptibility-induced signal loss, without compromising temporal resolution. It can be used for event-related functional experiments without requiring additional subject-specific calibration or calculation time.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Amygdala / blood supply
  • Artifacts
  • Brain Mapping
  • Calibration
  • Cerebral Cortex / blood supply
  • Echo-Planar Imaging / methods*
  • Humans
  • Imaging, Three-Dimensional
  • Magnetic Resonance Imaging / methods*
  • Oxygen / blood*
  • Oxygen Consumption / physiology
  • Signal Processing, Computer-Assisted*

Substances

  • Oxygen