J Sports Sci. 2010 May;28(7):733-42. doi: 10.1080/02640410903582776.

Validity and reliability of the Cortex MetaMax3B portable metabolic system.

Author information

1
Department of Physiology, Australian Institute of Sport, Belconnen, ACT, Australia.

Abstract

Portable indirect calorimetry systems offer the advantage of field-based measurements, but manufacturers rarely provide data about validity or reliability. In this study, we evaluated the validity and reliability of the Cortex MetaMax3B portable metabolic system. Validity was determined by comparing MetaMax3B results against those from a first-principles metabolic calibrator and an automated Douglas bag system. Reliability was obtained from duplicate exercise tests completed by eight athletes. Participants completed three identical incremental rowing tests on a Concept2 ergometer; two tests used the MetaMax3B and one test used the Douglas bag system. Compared with the metabolic calibrator, the MetaMax3B results were within 0.20 litres . min(-1) (7.8%) and 6.15 litres . min(-1) (4.0%) for [Vdot]O(2) and V(E), respectively. During exercise, the MetaMax3B results were within 0.16 litres . min(-1) (4.1%; [Vdot]O(2)), 0.32 litres . min(-1) (7.7%; [Vdot]CO(2)), and 3.22 litres . min(-1) (4.9%; V(E)) compared with the Douglas bag system. The MetaMax3B results were significantly higher for [Vdot]O(2) (P = 0.03) and [Vdot]CO(2) (P < 0.001). The typical error from duplicate exercise tests using the MetaMax3B ranged from 2.0% ([Vdot]O(2)) to 3.6% (V(E)). Our results show that the MetaMax3B provides reliable measurements of metabolic demand with adequate validity for field-based measurements.

PMID:
20419553
DOI:
10.1080/02640410903582776
[Indexed for MEDLINE]
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Loading ...
    Support Center