Format

Send to

Choose Destination
See comment in PubMed Commons below
Opt Lett. 2009 Nov 15;34(22):3556-8. doi: 10.1364/OL.34.003556.

Portable optical tissue flow oximeter based on diffuse correlation spectroscopy.

Author information

1
Center for Biomedical Engineering, University of Kentucky, Lexington, Kentucky 40506, USA.

Abstract

A portable diffuse correlation spectroscopy (DCS) flowmeter has been extended to measure both tissue blood flow and oxygenation (namely, DCS flow oximeter). For validation purposes, calf muscle blood oxygenation during cuff inflation and deflation was measured concurrently using the DCS flow oximeter and a commercial tissue oximeter. The oxygenation traces from the two measurements exhibited similar dynamic responses, and data were highly correlated (r(mean)>0.9, P<10(-5), n=10). The portable, inexpensive, and easy-to-use DCS flow oximeter holds promise for bedside monitoring of tissue blood flow and oxygenation in clinics.

PMID:
19927209
[Indexed for MEDLINE]
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for Optical Society of America
    Loading ...
    Support Center