Format

Send to

Choose Destination
See comment in PubMed Commons below
Opt Lett. 2009 Jul 15;34(14):2228-30. doi: 10.1364/OL.34.002228.

Scanning microscopy using a short-focal-length Fresnel zone plate.

Author information

1
School of Engineering and Applied Science, Harvard University, Cambridge, MA 02138, USA. schonbru@seas.harvard.edu

Abstract

We demonstrate a form of scanning microscopy using a short-focal-length Fresnel zone plate and a low-NA relay telescope. Owing to a focal length of only 5 microm, the zone plate produces large wavefront tilt and consequently severe vignetting for off-axis illumination. By scanning an optically trapped fluorescent sphere, we measure the vignetted collection region of the zone-plate imaging system. The fluorescence collection efficiency is sharply peaked and has a lateral width of 550 nm, which agrees with numerical simulations.

PMID:
19823557
PubMed Commons home

PubMed Commons

0 comments
How to join PubMed Commons

    Supplemental Content

    Full text links

    Icon for Optical Society of America
    Loading ...
    Support Center