Format

Send to

Choose Destination
J Vasc Interv Radiol. 2009 Sep;20(9):1188-92. doi: 10.1016/j.jvir.2009.05.033. Epub 2009 Jul 23.

Safety and effectiveness of the celect inferior vena cava filter: preliminary results.

Author information

1
Department of Radiology, Massachusetts General Hospital, USA.

Abstract

PURPOSE:

To evaluate the safety and effectiveness of the Celect inferior vena cava (IVC) filter during implantation, retrieval, and short-term follow-up.

MATERIALS AND METHODS:

The clinical data of 73 patients (46 men; age range, 22-89 years) who had a Celect IVC filter implanted between August 2007 and June 2008 were reviewed. Twenty-one (28.8%) presented with pulmonary embolism (PE), 15 (20.54%) with deep vein thrombosis (DVT), 12 (16.4%) with both, and the rest (34.26%) with other symptoms. Indications for filter placement were contraindication to anticoagulation (n = 38; 52%), prophylaxis/added protection (n = 22; 30%), failure of anticoagulation (n = 11; 15%), and complications of anticoagulation (n = 2; 3%). Filters were placed in the infrarenal (n = 71) or suprarenal (n = 2) IVC. Follow-up data were reviewed for filter-related complications and recurrent PE.

RESULTS:

All filters were successfully deployed. Immediately after fluoroscopy-guided filter deployment in 61 patients, four filters (6.5%) showed significant tilt. During follow-up (mean, 68 days +/- 73), three patients developed symptoms of PE after filter placement; however, computed tomographic (CT) pulmonary angiography demonstrated new PE in only two. Imaging follow-up with radiography (n = 32), CT (n = 11), and/or angiography (n = 4) in 47 patients (at a mean of 62 days +/- 75) showed no filter migration. Follow-up abdominal CT (at a mean of 69 days +/- 58) was available in 18 patients and demonstrated filter-related problems in seven (39%). These included penetration of filter legs in four and fracture/migration of filter components in one. Fourteen filters were successfully retrieved after a median period of 84 days.

CONCLUSIONS:

The Celect IVC filter can be safely placed but is related to a high incidence of caval filter leg penetration. Symptomatic PE after filter placement confirmed by CT occurred in 2.8% of patients.

PMID:
19631561
DOI:
10.1016/j.jvir.2009.05.033
[Indexed for MEDLINE]