Format

Send to

Choose Destination
Clin Implant Dent Relat Res. 2010 Sep;12(3):235-43. doi: 10.1111/j.1708-8208.2009.00156.x. Epub 2009 Mar 31.

Reliability and failure modes of implant-supported Y-TZP and MCR three-unit bridges.

Author information

1
Department of Prosthodontics, Integrated Center for Research, University of São Paulo, Bauru, SP, Brazil. estevamab@gmail.com

Abstract

PURPOSE:

Chipping within veneering porcelain has resulted in high clinical failure rates for implant-supported zirconia (yttria-tetragonal zirconia polycrystals [Y-TZP]) bridges. This study evaluated the reliability and failure modes of mouth-motion step-stress fatigued implant-supported Y-TZP versus palladium-silver alloy (PdAg) three-unit bridges.

MATERIALS AND METHODS:

Implant-abutment replicas were embedded in polymethylmethacrylate resin. Y-TZP and PdAg frameworks, of similar design (n = 21 each), were fabricated, veneered, cemented (n = 3 each), and Hertzian contact-tested to obtain ultimate failure load. In each framework group, 18 specimens were distributed across three step-stress profiles and mouth-motion cyclically loaded according to the profile on the lingual slope of the buccal cusp of the pontic.

RESULTS:

PdAg failures included competing flexural cracking at abutment and/or connector area and chipping, whereas Y-TZP presented predominantly cohesive failure within veneering porcelain. Including all failure modes, the reliability (two-sided at 90% confidence intervals) for a "mission" of 50,000 and 100,000 cycles at 300 N load was determined (Alta Pro, Reliasoft, Tucson, AZ, USA). No difference in reliability was observed between groups for a mission of 50,000. Reliability remained unchanged for a mission of 100,000 for PdAg, but significantly decreased for Y-TZP.

CONCLUSIONS:

Higher reliability was found for PdAg for a mission of 100,000 cycles at 300 N. Failure modes differed between materials.

[Indexed for MEDLINE]

Supplemental Content

Full text links

Icon for Wiley
Loading ...
Support Center