Rev Sci Instrum. 2009 Apr;80(4):043102. doi: 10.1063/1.3117208.

Use of ruthenium dyes for subnanosecond detector fidelity testing in real time transient absorption.

Author information

1
CEA, IBITECS, Laboratoire de Photocatalyse et Biohydrogène, F-91191 Gif-sur-Yvette, France. martin.byrdin@cea.fr

Abstract

Transient absorption spectroscopy is a powerful tool for the study of photoreactions on time scales from femtoseconds to seconds. Typically, reactions slower than approximately 1 ns are recorded by the "classical" technique; the reaction is triggered by an excitation flash, and absorption changes accompanying the reaction are recorded in real time using a continuous monitoring light beam and a detection system with sufficiently fast response. The pico- and femtosecond region can be accessed by the more recent "pump-probe" technique, which circumvents the difficulties of real time detection on a subnanosecond time scale. This is paid for by accumulation of an excessively large number of shots to sample the reaction kinetics. Hence, it is of interest to extend the classical real time technique as far as possible to the subnanosecond range. In order to identify and minimize detection artifacts common on a subnanosecond scale, like overshoot, ringing, and signal reflections, rigorous testing is required of how the detection system responds to fast changes of the monitoring light intensity. Here, we introduce a novel method to create standard signals for detector fidelity testing on a time scale from a few picoseconds to tens of nanoseconds. The signals result from polarized measurements of absorption changes upon excitation of ruthenium complexes {[Ru(bpy)(3)](2+) and a less symmetric derivative} by a short laser flash. Two types of signals can be created depending on the polarization of the monitoring light with respect to that of the excitation flash: a fast steplike bleaching at magic angle and a monoexponentially decaying bleaching for parallel polarizations. The lifetime of the decay can be easily varied via temperature and viscosity of the solvent. The method is applied to test the performance of a newly developed real time transient absorption setup with 300 ps time resolution and high sensitivity.

PMID:
19405646
DOI:
10.1063/1.3117208

Supplemental Content

Full text links

Icon for American Institute of Physics
Loading ...
Support Center